Inlet systems
The inlet system provides the means of introducing the specimen into the GC. Obtaining a narrow sample band at the start of the chromatographic process is critical to achieve good resolution, since broad sample bands usually produce broad peaks, especially for analytes that elute early. The choice of injector depends on:
• the characteristics of the specimen or residue
• the quantity and characteristics of the analytes to be separated
• the temperature and nature of the stationary phase and the column.
Solid samples are normally chromatographed by dissolving in a suitable solvent and injecting with a micro-syringe. Liquids can be injected using a micro-syringe, but with sensitive detection systems the sample should be dissolved in a suitable solvent to reduce the sample size and avoid overloading the detector. Gases and vapours may be introduced by injection through the inlet port septum using a gas-tight syringe. The four main types of GC injectors are mega bore direct (or packed column), split, splitless and cold on-column. In reality, split and splitless injection are carried out using the same hardware; the so-called split/splitless injector. Conventional glass syringes of 1–10 μL volume with stainless-steel needles can be used on the packed column, or vaporisation capillary injectors, and the injection is made by piercing a silicone rubber septum. Care must be taken to select septa that have low bleed characteristics at the operating temperature, and those with Teflon backs are most reliable in this respect. Unstable solutes can be decomposed by the high temperature of the injection system, particularly if the system is constructed of metal. For labile substances, cold on-column injection is preferred, but clean extracts must be used to minimise column contamination. The injection system typically utilises a removable liner to minimise contact with metal surfaces. These liners are usually made of glass. They come in various configurations and the type used depends on the nature of the sample being analysed, the injector system used and the injection volume. With repeated injection of samples, liners inevitably become contaminated with nonvolatile material. This can cause unwanted interactions with subsequent injection of samples. Hence liners can be removed, cleaned by solvent washing and replaced in the injector.