Installing, conditioning and maintaining columns
Temperature programming
For complex mixtures with components of widely varying retention characteristics, it is often impractical to choose a single column temperature that allows all the components to be resolved. Increasing the column temperature throughout the analysis dramatically reduces the time taken for higher-boiling compounds to elute, and simultaneously improves the sensitivity of the assay, as the peaks are remarkably sharper. If the early-eluting compounds are resolved inadequately, a lower starting tempera ture or slower initial ramp should be used, taking care to observe the temperature requirements of the type of injector used. All instruments currently manufactured are available with a temperature program option, and a multi-ramp programmer is particularly useful for capillary chromatography. The first ramp can be used during splitless injection to bring the column rapidly up to the initial chromatography temperature, followed by a slower analytical ramp to perform the separation. One problem with temperature programming is that the back pressure increases with temperature and reduces the carrier gas flow if a mass-flow controller is not used. For polar stationary phases, the polarity increases with temperature. Column bleed also increases, which results in an increasing baseline. For this reason the column should be well conditioned