Evaluating column performance
Retention time of a non-retained compound or hold-up time (tM)
The retention time tM is the time taken for a non-retained solute to travel along the column; it represents the transit time for the mobile phase (carrier gas) in the column and is a column specific parameter, applicable only under the prevailing conditions of gas flow and oven temperature. No other peak can be expected to elute earlier than this time. tM is obtained by injecting a non-retained compound suitable for the detector system being used (butane or methane for flame ionisation detection (FID) or thermal conductivity detection (TCD); acetonitrile for nitrogen–phosphorus detection (NPD); methylene chloride for electron-capture detection (ECD); vinyl chloride for photoionisation detection (PID) or electrolytic conductivity detection (ELCD)).